Wuhan Goldhome Hipot Electrical Co., Ltd. is one of the leading manufacturers and suppliers of automatic water-soluble acid test equipment in China. With abundant experience, our factory offer high quality customized products made in China with competitive price. Welcome to contact us for quotation.
Products Video
HM203B Fully Automatic Water-soluble Acid Tester Description Video
Product Overview

The design and manufacture of the HM203B Fully Automatic Water-Soluble Acid Tester (Colorimetric Method) comply with the People's Republic of China National Standard GB/T7598-2008 "Test Method for Water-Soluble Acids in Transformer Oil During Operation" and related standards. Featuring a large LCD display and microprocessor control, it automatically performs heating, temperature maintenance, agitation, sampling, testing, cleaning, and printing according to user-set parameters. Capable of simultaneously measuring 1 to 6 samples, it significantly reduces testing time.
Features & Advantages

1. This instrument is designed and developed in accordance with GB/T7598-2008 standards, serving as a fully automated testing device for accurately detecting water-soluble acids in petroleum products such as transformer oil, turbine oil, and fire-resistant oil.
2. Features a color touchscreen LCD for intuitive operation via touch-sensitive display.
3. Utilizes colorimetry to determine pH values.
4. The instrument features high automation. Simply inject oil and water according to standard specifications, and the device automatically performs heating, agitation, settling, oil-water separation, color development, colorimetry, result display, and printing.
5. It can process and sequentially complete testing for 1 to 6 samples at once.
6. The integrated design offers a more aesthetically pleasing appearance; the large 7-inch screen display simplifies operation.
Technical Specifications
|
Parameter |
Specifications |
| Measuring range | PH=3.6 ~ 7.0 |
| Minimum display unit | 0.1 |
| Constant temperature accuracy: | ±2℃ |
| Ambient temperature | (5 ~ 45) ℃ |
| Relative humidity | ≤ 85% |
| Power supply | AC (220± 10%) V, 50HZ±1HZ |
| Dimensions | length 660 mm; Width 415mm; Height 470 mm |
| Weight: | 36kg |
Instructions for use
1. Preparation process before starting
(1) Measure 50ml of the oil sample to be measured into 250ml conical bottle, add 50ml of pre-boiled distilled water, cover the matching bottle stopper of the conical bottle and put it into the corresponding slot of the instrument, connect the corresponding connecting pipe, clamp it with a fixed rod, and finally cover the box.
(2) Fix the indicator (behind the instrument), put the bromocresol green and bromocresol purple (with the corresponding label) indicators into the corresponding position, and connect the corresponding connection pipe.
(3)Connect the instrument to the power supply, grounding the instrument shell reliably, and turn on the power switch.
(4 )Instrument operation process: The boot interface is shown in Figure 1

Figure 1
The default national standard parameters of the instrument, such as direct selection of quick test without changing, select the cup position and directly click the start key (Figure 2), you can start the test.
Figure 2
After starting, enter the test interface (Figure 3), and automatically complete the test process of heating, shaking, static, testing, draining, printing data, etc. The test can be finished at any time in this interface.
Note: In automatic test, the instrument first uses bromocresol violet as indicator. If the test fails, the buzzer will warn to change the cup. After changing the test cup, press the confirm button to continue the determination with bromocresol green.
Figure 3
Press the system setting key to enter the system setting interface (Figure 4), which can be set: oscillation time, rest time, heating temperature, printing mode.
Figure 4
Oscillation time setting interface (Figure 5), set the required proper time, default 300S. After setting, press the back key to save the set time automatically.
Figure 5
Static time setting interface (Figure 6), set the required proper time, default 600S. After setting, press the back key to save the set time automatically.
Figure 6
Heating temperature setting interface (Figure 7), set the required proper time, the default 75℃. After setting, press the back button to automatically save the set temperature.
Figure 7
Select the printing method (Figure 8), you can choose to print automatically after the completion of the test or manually print by yourself.

Figure 8
System maintenance interface (Figure 9), system pipeline cleaning and exhaust, clock time setting.
Figure 9
The pipe cleaning interface (Figure 10) can manually clean the residual liquid in the 1-6 cup pipe and indicator pipe. If there is oil sample in the pipeline, it is necessary to clean with anhydrous ethanol. Note: The excess air in the pipeline must be emptied to avoid affecting the amount of liquid added and affecting the experimental results.
Figure 10
Time setting interface (Figure 11), you can set the system clock time.
Figure 11
The data query interface (Figure 12) can query the historical data made and print it.
Figure 12